Minolta Pulsox-2 Instruction Manual Page 23

  • Download
  • Add to my manuals
  • Print
  • Page
    / 24
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 22
21
Guidance and manufactures declaration - electromagnetic immunity
The Model PULSOX-2 is intended for use in the electromagnetic environment specified below. The
customer or the user of the Model PULSOX-2 should assure that it is used in such an environment.
Immunity test IEC 60601
Test level
Compliance
level
Electromagnetic environment -
guidance
Radiated RF
IEC61000-4-3
3 V/m
80MHz to 2.5GHz
Conducted RF
IEC61000-4-6
Not applicable
3 V/m
NOTE 1 At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephone and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the PULSOX-2 is used exceeds the
applicable RF compliance level above, the PULSOX-2 should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating the PULSOX-2.
b
Over the frequency range 150kHz to 80MHz, field strengths should be less than 3V/m.
Portable and mobile RF communications
equipment should be used no closer to
any part of the PULSOX-2, than the
recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.

80MHz to 800MHz
800MHz to 2.5GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the recommended
separation distance inmeters (m).
Field strength from fixed RF transmitters,
as determined by an electromagnetic site
survey
a
, should be less than the
compliance level in each frequency
range
b
.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
Recommended separation distance
E1
E1
9222-1731-14_P-2.indd Sec1:219222-1731-14_P-2.indd Sec1:21 09.8.20 11:03:24 AM09.8.20 11:03:24 AM
Page view 22
1 2 ... 18 19 20 21 22 23 24

Comments to this Manuals

No comments